Le Lézard
Classified in: Health
Subject: RCL

Smiths Medical Issues Urgent Medical Device Correction Letter Notifying Customers of Potential Issues with CADDtm Infusion System Infusion Sets for use with CADD pumps


MINNEAPOLIS, Dec. 12, 2022 /PRNewswire/ -- Smiths Medical issued an Urgent Medical Device Correction Letter to notify affected customers of two potential issues with CADDTM Infusion System Infusion Sets related to potential lack of delivery or underdelivery and false no disposable attached (NDA) alarms. The letter details the issues, the affected items, the required steps to perform, and specific instructions for treatment of patients requiring life sustaining therapy.

The first issue, lack of delivery or underdelivery, may occur due to manufacturing variations that can potentially cause the green CADD Flow Stop arm to compress partially occlude the tubing before clinical use. If this happens, there is a potential that the occlusion does not resolve when the CADD reservoir or administration set is connected to the pump, and the pump may not detect the occlusion. This may result in underdelivery or non-delivery of medication, despite the pump displaying that the infusion is running properly.

The second issue, false "no disposable attached" (NDA) alarms, is specific to CADD-Legacy pumps, which Smiths Medical announced the discontinuation of sale effective December 31, 2022. There is a potential that CADD-Legacy pumps may not detect that 50 mL and 100 mL CADD Medication Cassette Reservoirs with Flow Stop are attached to the pump when they are properly attached. When this happens, the pump will initiate an NDA alarm if the NDA double-beep warning is not resolved within 2 minutes. The user must clear the alarm and resolve the cause of the NDA event before using the pump. This issue does not impact 250 mL Flow-Stop and non-Flow Stop CADD Medication Cassette Reservoirs.

A copy of the customer notification, which outlines the potential risks associated with both underdelivery and NDA alarms, lists the affected products, and provides specific actions users should take in the event of such occurrences, can be found here.

For further inquiries, please contact Smith Medical using the information provided below.

Smiths Medical Contact                   

Contact Information                                 

Areas of Support

Global Complaint Management 

[email protected]   

1-(866)-216-8806                                       

To report adverse events or
product complaints

Technical Assistance                       

1-(800)-258-5361                             

Additional information or
technical assistance

The U.S. Food and Drug Administration (FDA) has been notified of this action.

If you wish to contact the FDA regarding any adverse events or quality problems associated with this notice, use the following contact information.

Media Contact: 
Tom McCall 
949-366-4368 
[email protected]

SOURCE ICU Medical, Inc.


These press releases may also interest you

at 05:02
'Companion Diagnostics: Technologies and Markets' is all about using special tests alongside certain medicines to figure out which treatment will work best for each person. These tests are called companion diagnostics because they go together with...

at 05:00
WellSync, an API-driven virtual care service provider, today announced its partnership with The Vitamin Shoppe to launch Whole Health Rx, a cutting-edge telehealth service providing access to GLP-1 medications for weight management. WellSync provides...

at 04:00
May 17, 2024 Three studies demonstrate how Philips MCOT wearable ambulatory monitoring ECG and proprietary AI models applied to ECG digital biomarkers can help to improve diagnosis, reduce readmissions, and lower costs Amsterdam, the Netherlands...

at 04:00
Brainomix, a pioneer in artificial intelligence (AI) imaging solutions to enable precision medicine, is proud to announce the FDA clearance of...

at 04:00
OMRON Healthcare Co., Ltd. based in Muko, Kyoto Prefecture, Japan, announced its presence at Heart Rhythm 2024, the Annual Meeting of the Heart Rhythm Society (HRS) in the U.S. At Heart Rhythm 2024, the Company will showcase its upper-arm blood...

at 03:00
Hypertension, or high blood pressure, affects one in three Europeans over the age of 30 and more than one billion people worldwide.  Hypertension is known as the "silent killer" because many people do not know they have it and, over time, it can lead...



News published on and distributed by: