Smiths Medical Issues Urgent Medical Device Correction Letter Notifying Customers of a Potential Issue with Level 1® H-2 Pressure Chambers used with Level 1 Fast Fluid Flow Fluid Warmers


MINNEAPOLIS, Aug. 3, 2022 /PRNewswire/ -- Smiths Medical issued an Urgent Medical Device Correction Letter to notify affected customers of a potential issue with specific Level 1® H-2 Pressure Chambers used with the Level 1 Fast Flow Fluid Warmers.

In 2015, Smiths Medical implemented a design change to widen the hinge/latch assembly on the Level 1 H-2 Pressure Chambers used with the Level 1 Fast Flow Fluid Warmers (Models H-1025 or H-1200) or added it to the H-1000 model. The company has become aware that Level 1 H-2 Pressure Chambers with the wider hinge/latch assembly can potentially impact the pressure exerted onto the IV fluid bag while contained within the pressure chamber, which may result in decreased flow rate, stopped flow, or residual fluid left within the IV bag.

Affected product was distributed in the United States between December 19, 2016, and March 10, 2022.  However, all Level 1 H-2 Pressure Chamber devices have the potential to be affected by this issue, because some devices may have received a hinge/latch replacement during that timeframe. Level 1 H-2 Pressure chambers with the wide hinge/latch assembly are more susceptible to this issue in the following scenarios:

  1. kinked tubing on the disposable administration sets
  2. use of the lowest flow rate disposables (DI-50, D-70, or DI-70) when delivering viscous fluids such as chilled blood from 300 mL or smaller IV bags

Decreased flow rate, stopped flow, or residual fluid left within the IV bag could result in under-delivery or delay of therapy, leading to potential inadvertent hypothermia, hypovolemia, and/or hypotension which may lead to serious injury and death.

Smiths Medical has sent all affected customers and distributors a letter outlining the risk and providing specific steps to determine whether their devices are affected or not and what steps to follow if they have affected product in service. A copy of that letter can be found here

For further inquiries, please contact Smith Medical using the information provided below.

Smiths Medical Contact

Contact Information

Areas of Support

Global Complaint Management

[email protected]

1-(866)-216-8806

To report adverse events or product complaints

Device Correction Inquiries

1-(800)-241-4002, option 4, then prompt 1

For any questions regarding this action

Technical Assistance

1-(800)-258-5361, option 2, then prompt 6

Additional information or technical assistance

The U.S. Food and Drug Administration (FDA) has been notified of this action.

Adverse reactions or quality problems experienced with this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.

Media Contact:
Tom McCall
949-366-4368
[email protected]

SOURCE Smiths Medical


These press releases may also interest you

at 05:35
Drug Farm announced today the U.S. Food and Drug Administration (FDA) has cleared the Investigational New Drug (IND) application of DF-003, a first-in-class, oral, potent, highly selective alpha-kinase 1 (ALPK1) inhibitor for clinical evaluation in...

at 05:02
'Companion Diagnostics: Technologies and Markets' is all about using special tests alongside certain medicines to figure out which treatment will work best for each person. These tests are called companion diagnostics because they go together with...

at 05:00
WellSync, an API-driven virtual care service provider, today announced its partnership with The Vitamin Shoppe to launch Whole Health Rx, a cutting-edge telehealth service providing access to GLP-1 medications for weight management. WellSync provides...

at 04:00
May 17, 2024 Three studies demonstrate how Philips MCOT wearable ambulatory monitoring ECG and proprietary AI models applied to ECG digital biomarkers can help to improve diagnosis, reduce readmissions, and lower costs Amsterdam, the Netherlands...

at 04:00
Brainomix, a pioneer in artificial intelligence (AI) imaging solutions to enable precision medicine, is proud to announce the FDA clearance of...

at 04:00
OMRON Healthcare Co., Ltd. based in Muko, Kyoto Prefecture, Japan, announced its presence at Heart Rhythm 2024, the Annual Meeting of the Heart Rhythm Society (HRS) in the U.S. At Heart Rhythm 2024, the Company will showcase its upper-arm blood...



News published on and distributed by: